Meinecke Resource

Meinecke’s factory calibration certificates can be a bit difficult to decipher the first time you see one, so we have provided the sample certificate along with a brief explanation below of the typical data found on these certificates.
They are written in 3 languages, English is usually the 2nd one listed. The meters are tested in the factory’s calibration shop under real conditions. We’ll translate the first few lines.
They are written in 3 languages, English is usually the 2nd one listed. The meters are tested in the factory’s calibration shop under real conditions. We’ll translate the first few lines.
- Type: WPD 100 = 10mm or 4 inch WP/Dynamic Flow meter
- For: 50’c and PN:16 = Rated to 122’F and Pressure Rated to 16 Bar
- Order No. – This is the Sales Order we as a Master Distributor purchased these meters on.
- Meter No = Serial Number, Flow = Rate in M3/Hr, Date Tested: 16.02.07 = February 16, 2007
In the main body, you’ll see all the Serial Numbers of meters tested in that group in the first column. Meters are gang tested so you may see more than one meter listed on your report.
Meinecke tests these meters at three different relevant flow rates. For your 3” meter, test flow rates are shown below along with the deviation from that rate by %..

The next 3 columns show the deviation by % from the standard reading, at three distinct flow rates.
- 200 m3/h (660.57 GPM) / -1.0% deviation
- 1.8 m3/h (7.92 GPM) / +1.2% deviation
- 0.8 m3/h (3.52 GPM) / -0.5% deviation
The deviations listed are from the first meter listed… Serial Number: 07 0016987
Are you interesting in learning more? Fill out our contact form with some details of your application and we’ll have a specialist get in touch! Also, learn more about Meinecke as a products and solutions provider for Cross.